Temperature Instrumentation & Calibration Guide

Technical Articles

Home/ Service & Support/ Technical Articles/ Temperature Instrumentation & Calibration Guide
Author: admin-管理员
Temperature instrumentation calibration: probes in a dry-block calibrator wired to a digital indicator

Pyrometry & Thermocouples

Thermocouple type selection, compensating cable, installation checklist and AMS 2750 under SAT/TUS test procedure.
Industrial-furnace temperature-measurement system: thermocouple type and selection, compensating cable, installation standard,AMS 2750 under the framework of SAT system-accuracy test and TUS furnace-temperature uniformity survey procedure.

Thermocouple type

Type Calibration Range Features
K type ( Ni-Cr – nickel-silicon) NiCr-NiAl −200–1260°C Most common, low cost, oxidation resistance is moderate, long-term drift
N type Nicrosil-Nisil −200–1300°C Ratio K type oxidation-resistant, better stability, high-temperature furnace trend
R/S type ( platinum-rhodium) PtRh-Pt 0–1600°C Noble metal, high accuracy, used for instrument and TUS
B type ( platinum-rhodium 30- platinum-rhodium 6) Pt30Rh-Pt6Rh 0–1800°C High temperature, no thermoelectric EMF at ambient

Compensating cable

Must match the thermocouple calibration (K type uses green +/−, extension-wire type KX); polarity must not be reversed; single-point ground to avoid ground loops; in high-temperature zones use heat-resistant extension wire or extend the thermocouple itself.

Thermocouple installation checklist

  • Insertion depth ≥ protection-tube inner diameter 8–10 times, place the hot junction at a representative part location;
  • Avoid contact with the furnace wall / heating element, ceramic tube free of cracks;
  • Terminal screws tight, extension-wire polarity correct;
  • Note the grounding method for mineral-insulated thermocouples ( insulation / grounded) match the controller;
  • Record number, location, insert date and reading.

SAT system-accuracy test (AMS 2750)

  1. Use a standard thermocouple (R/S type) compare at the same location as the instrument thermocouple;
  2. Take multiple points above and below the working temperature ( such as 3 point) measurement;
  3. System error = instrument reading − standard value, should be within tolerance (Class 2:±3°C or 0.3%);
  4. On out-of-tolerance, check the extension wire, terminal, cold-junction compensation and instrument calibration.

TUS furnace-temperature uniformity test

  • Loaded thermocouple ≥ 9 pcs (Class 2 furnace), arrange to cover the work zone;
  • Empty furnace / full load must all be tested, perform per the calibration cycle ( usually 6–12 months);
  • Acceptance band ±6°C(Class 2), if exceeded, adjust combustion / power zoning or shorten the soaking window;
  • Archive the test data, bound to the heat-treat record.
This site’s pyrometry and instrumentation references are companion tools for optimizing the temperature-control system.
Related tools:PT100 conversion
Technical Reference Center · Compiled from industrial furnace engineering practice and industry standards.

Instrumentation & Measurement

Temperature / pressure / flow / level-instrument calibration procedure, record and uncertainty evaluation

Professional guide to furnace-instrument calibration: temperature, pressure, flow, calibration basis for level instruments, equipment configuration, operating procedure, record elements and uncertainty evaluation. Based on ISO 9001,ISO/IEC 17025,AMS 2750 and IEC 60584 and other standard references, suitable for daily furnace-instrument management and quality-system building.
Professional-reference disclaimer:This page is a general reference to industry standards and best practices. For critical applications, verify the current valid standard text (ISO/BS/ASTM/ASME) against the equipment manufacturer manual; cost data are UK-market approximations; safety-critical work must be performed by certified personnel per regulations. Content current as of 2025 year 2 month.

1. Calibration basis and standards

Calibration= compare instrument readings with known reference standards, then adjust or record the error.

  • Accuracy:Ensure correct measurement ( quality, safety, key to legal compliance)
  • Traceability:Measurement results traceable to the national standard ( UK NPL, US NIST)
  • Regulatory requirement:ISO 9001,AMS 2750( heat treatment),FDA( pharmaceutical) all require instruments under controlled calibration

1.1 calibration interval

Instrument type Typical interval Cases for shortening the interval
Process thermocouple 6-12 months Harsh environment (>1000°C), frequent thermal cycling
RTD(PT100) 12 months Vibration, corrosive atmosphere;AMS 2750 requirement 6 months
Pressure transmitter 12-24 months Overpressure event, fouling / corrosive medium
Flowmeter 12-36 months Abrasive fluid, custody transfer ( shortened to 6-12 months)

1.2 external calibration (UKAS/ISO 17025) compared with internal calibration

Method Executor Cost ( UK reference) Applicable occasions
External calibration by an accredited laboratory ISO 17025 accredited laboratory ( traceable to the national standard) £80-300 / pcs Legal metrology,AMS 2750,FDA/ pharmaceutical, customer requirement
Internal calibration A competent technician uses accredited-calibrated reference standards £10-50( manual only) Internal QC,ISO 9001( when the procedure is documented)

Key rules:The reference standard itself must be calibrated by an accredited body; process instruments can be calibrated internally using these reference standards.

2. Temperature-instrument calibration

2.1 equipment requirement

Equipment Typical specification Cost ( UK reference)
Dry-well calibrator Range 50-660°C, stability ±0.1°C, accuracy ±0.25°C £1,500-5,000
Reference thermocouple K type 1 class or SLE, accredited calibration £150-400
Precision thermometer Resolution 0.01°C, accredited calibration £500-2,000

2.2 thermocouple calibration procedure

  1. Installation:Insert the reference and unit-under-test thermocouples together into the dry-well holes ( improve contact with thermal grease)
  2. Set:Start from the lowest calibration point ( such as 200°C)
  3. Stable:Wait 10-15 min (±0.5°C maintain 5 min)
  4. Record:Reference = 200.2°C, unit under test = 199.5°C → error = −0.7°C
  5. Multi-point:Take 3-5 points ( such as 200/400/600/800/1000°C)
  6. Documented:All readings and errors recorded in the calibration certificate
  7. Adjust:When error exceeds tolerance, adjust the controller offset or apply a correction

2.3 thermocouple tolerance (IEC 60584)

Model 1 class tolerance 2 class tolerance Range
K type ±1.5°C or ±0.004×T ±2.5°C or ±0.0075×T −40 ~ +1200°C
N type ±1.5°C or ±0.004×T ±2.5°C or ±0.0075×T −40 ~ +1300°C
R type ±1.0°C or ±[1+0.003×(T−1100)] ±1.5°C or ±0.0025×T 0 ~ +1600°C

2.4 PT100 resistance – temperature comparison

Temperature Resistance (Ω) A class tolerance
−50°C 80.31 ±0.24°C
0°C 100.00 ±0.15°C
100°C 138.51 ±0.35°C
200°C 175.86 ±0.55°C
400°C 247.09 ±0.95°C

3. Pressure-instrument calibration

3.1 equipment

Equipment Range Accuracy Cost ( UK reference)
Piston gauge (deadweight tester) 0-700 bar ±0.015%( optimal) £3,000-15,000
Digital pressure calibrator 0-400 bar ±0.025-0.05% FS £2,000-8,000
Hand pump + standard gauge 0-40 bar ±0.1% FS £500-2,000

3.2 calibration procedure

  1. Connection:Hand pump → reference gauge → transmitter under test ( in series, via the valve manifold)
  2. Zero point:Vented to atmosphere (0 bar) should read 4.00mA
  3. Pressurize:Of the range 25% / 50% / 75% / 100%
  4. Record:At each point record the reference-gauge reading and transmitter output
  5. Reduce pressure:Reduce pressure in the reverse direction to check hysteresis
  6. Calculate:The maximum error should < ±0.5% FS
  7. Adjust:Zero point / range screw or HART communicator

3.3 4-20mA signal calculation

Formula:MA = 4 + 16 × range percentage
Example:0-10 bar transmitter, actual pressure 7.5 bar → range 75% → expected 16.00mA; if measured 16.3mA → error +0.3mA = +1.875% FS( out of tolerance).

3.4 pressure-gauge accuracy class (BS EN 837-1)

Class Accuracy (% FS) Typical use
0.1 class ±0.1% Reference standard
0.25 class ±0.25% Verification gauge
1.0 class ±1.0% Process gauge ( industrial standard)
1.6 class ±1.6% General indication

4. Flowmeter calibration

4.1 calibration methods for each flowmeter type

Type Method Accuracy Cost ( UK reference)
Electromagnetic flowmeter Dry type ( electronic) or wet type ( weighing) ±0.2-0.5% £80-2,000
Coriolis mass flowmeter Electronic verification or wet type ±0.1-0.2% £150-5,000
Turbine flowmeter Wet calibration only ±0.5-1.0% £500-3,000
Orifice plate / differential pressure Measure the orifice diameter + calibrate the differential-pressure transmitter ±1-3% £150-500

4.2 calibration by weighing method ( gold standard for liquid flow)

  1. Switch the flow to the collection vessel on the weigh platform
  2. Start the timer and record the initial weight
  3. Collect a known time interval ( such as 60 s)
  4. Stop and record the final weight
  5. Calculate: flow =( weight / density)/ time
  6. Compared with the flowmeter reading, adjust K coefficient

4.3 orifice-plate inspection

  • Orifice diameter: caliper measurement (±0.01mm), compared with the nameplate
  • Inlet edge: must be sharp ( free of burrs, roundness <0.0005d)
  • Flatness: check flatness with a straightedge
  • Pressure tap: clean and unobstructed, correct spacing (D and D/2 or flange tapping)
  • Replacement condition: orifice diameter deviates from nominal >±0.5%, edge damage, visible corrosion

5. Level-instrument calibration

5.1 type and method

Type Principle Method Accuracy
Differential-pressure type Static pressure P = ρgh Calibrate the differential-pressure transmitter 0-100% ±0.5-1% FS
Ultrasonic Acoustic path time in air Compared with tape-measure field measurement ±1-3% FS
Radar ( guided-wave) Microwave time-of-flight Dry or wet calibration ±1-5 mm
Capacitive type Permittivity change Analog or wet calibration ±1-2% FS

5.2 differential-pressure level transmitter

Theory:P = ρgh
Example:Water tank 0-5m, water density 1000 kg/m³ → 5m pressure at = 1000 × 9.81 × 5 = 49,050 Pa = 0.49 bar → transmitter range 0-0.49 bar corresponding 4-20mA.

5.3 calibration steps

  1. Isolation:Close HP/LP valve, open the equalizing valve
  2. Connection:Hand pump + reference gauge connected to HP side,LP vented to atmosphere
  3. 0% calibration:Add 0 mbar should read 4.00mA
  4. 100% calibration:Add ρgh the corresponding pressure should read 20.00mA
  5. Linearity:Test 25/50/75%( increase pressure + reduce pressure)
  6. Adjust:Zero point / range button or HART
  7. Restore:Close the equalizing valve, open the isolation valve, leak detection

6. Calibration procedure and records

6.1 required record elements

  • Instrument identification: tag number, serial number, brand / model
  • Date + technician name
  • Reference standard: brand / model / serial number + its validity period ( must be within the calibration validity period)
  • Environment: temperature ±2°C, humidity, atmospheric pressure
  • As-Found data:Pre-adjustment reading
  • As-Left data:Post-adjustment reading
  • Pass / fail criterion ( based on the tolerance)
  • Next calibration date = calibration date + interval
  • Affix labels: green ” calibration due:MM/YYYY”

6.2 determine tolerance by application

Application Typical tolerance Reason
General indication ±1-2% FS Non-critical, monitoring only
Control loop (PID) ±0.5-1% FS Affects product quality
Safety interlock ±0.25-0.5% FS Failure implies a safety risk
AMS 2750 temperature measurement ±3°C or ±0.5% Statutory requirement ( aerospace)
Custody transfer ±0.1-0.25% Involving financial settlement

6.3 label-color management

Color Status Action
Green Calibration valid Normal use
Amber Approaching expiry (30 day(s)) Schedule calibration, may continue to be used
Red Overdue for calibration / fail Prohibited— re-calibrate

7. Measurement uncertainty and TUR

7.1 test uncertainty ratio (TUR) (TUR)

TUR = tolerance of the device under test / reference-standard uncertainty
Criteria:TUR ≥ 4:1( preferred), minimum 3:1(ISO 17025).
Example:Process-transmitter tolerance ±0.5% FS = ±0.05 bar(0-10bar); reference-gauge uncertainty ±0.015 bar → TUR = 3.3:1 → acceptable.

7.2 uncertainty source

Source Contribution Reduction measures
Reference-standard accuracy 50-70% Use a better reference ( piston type vs digital)
Resolution 10-20% Improve resolution (0.01°C vs 0.1°C)
Temperature effect 5-15% Controlled environment (20°C ±2°C)
Repeatability 5-10% Multiple readings (3-5 times) take the average

7.3 combined uncertainty

Root-sum-square:U = √(U₁² + U₂² + U₃² + …)

8. Temperature controller and instrument configuration

Common furnace temperature-control instruments include single-loop PID controller ( such as Eurotherm 3216/3504 series), multi-loop program controller ( heat-treatment process curve), recorder and over-temperature protector. Configuration key points:

  • Input type must match the sensor (K/N/R/S/B/PT100), cold-junction compensation enabled
  • Output type selected by load ( relay /SSR drive /4-20mA/ thyristor (SCR))
  • Over-temperature protection must useIndependent instrument, do not use PID controller software alarm substitution
  • The calibration interval for the controller and safety instruments is per AMS 2750 or executed under an internal quality system
  • Communication (Modbus RTU/RS-485) used for data acquisition and audit trail

See the terminal and output configuration for detailsWiring reference; see furnace temperature uniformity testing.

9. Oxygen probe ( carbon-potential sensor)

The oxygen probe measures the atmosphere oxygen potential via a zirconia element, convert to carbon potential (CO/CO₂ balance). Maintenance key points:

  • Reference gas ( clean air) stable flow, unobstructed piping
  • Install the probe away from direct impingement of the carburizing-agent nozzle, sense a representative atmosphere
  • Periodic carbon burn-off cleaning (soot cleaning) checked against impedance
  • Use standard gas per the manufacturer interval ( such as 1% O₂) verify the millivolt output
  • Probe at end of life ( usually 12-24 months) then output attenuation, replace promptly

See carbon-potential calculation.

10. Vacuum gauge

Type Range Depends on gas type Description
Bourdon tube / diaphragm Ambient air ~ approx. 1 mbar No Mechanical direct measurement
Thermocouple vacuum gauge Ambient air ~ 10⁻³ mbar Is Resistance vacuum gauge, low cost
Pirani gauge Ambient air ~ 10⁻⁴ mbar Is Heat-conduction type
Capacitance diaphragm gauge (CDG) Ambient air ~ 10⁻⁵ mbar No Independent of gas type, high accuracy
Ion gauge ( Penning / cold cathode) 10⁻³ ~ 10⁻⁹ mbar Is High-vacuum measurement

Calibration: thermocouple / the ion gauge can be compared with a capacitance diaphragm gauge; gauge-tube contamination ( oil vapor, dust) causes reading drift, clean or replace periodically.

11. CQI-9 heat-treatment system assessment

CQI-9(AIAG special process: heat-treatment system assessment) is a mandatory requirement of the automotive supply chain, core concern:

  • High-temperature test (Pyrometry):TUS/SAT execute, instrument calibration, record retention ( according to AMS 2750 or equivalent)
  • Process:Process specification, process monitoring ( temperature / carbon potential / time), mistake-proofing
  • Equipment capability:Furnace temperature uniformity, quench system, atmosphere control
  • Personnel and records:Training matrix, audit trail, nonconforming-product control

12. AMS 2750 high-temperature measurement

AMS 2750″ high-temperature measurement ” is the baseline standard for aerospace heat treatment, requirements for furnace instruments:

Item Requirement
Furnace temperature class Classify per process requirements (Class 1-6), uniformity ±3°C ~ ±15°C not equal
Instrument calibration Thermocouple and RTD calibration interval 6 months; reference standard traceable
TUS Before first use + routine interval ( typically every 6 months)+ after major overhaul
SAT Quarterly ( maximum 3 months)+ before commissioning after repair
Record retention Calibration certificate,TUS/SAT report, over-temperature record retention period per contract / specification

Related tools:·

Related tools:Wiring reference
Technical reference center · content compiled with reference to industrial-furnace engineering practice and industry standards

Comments

No comments yet. Feel free to ask your model-selection or technical questions; we will reply after review.

Comments are closed on this page.

Dingluxuan Sensor Technology

Design, manufacturing and sales of temperature, pressure and humidity sensors, instruments and ceramic products — delivering integrated temperature-measurement and control solutions.

Contact Us

Contact
No. 828 Xueyuan Road, Bacheng Town, Kunshan City, Jiangsu, China
Tel
0512-57892201
Mobile: 13451695112
Email: yanghui@dingluxuan.com
Copyright©2017-2026  Suzhou Dingluxuan Sensor Technology Co., Ltd.   苏ICP备17068273号-1