Correct installation is essential for accurate, reliable temperature measurement. A £50 thermocouple installed wrongly is worse than a £500 one installed correctly.

immersion depth

Rule of thumb: minimum immersion depth = 10 × sheath diameter.6mm sheathat leastinsertionprocess 60mm.

Issue Cause Fix
heat shunting(conductionerror) Heat conducts along the thermocouple sheath to the cold installation point, so the sensor reads low; the shallower the immersion, the larger the error. Increase the immersion depth; use a low-conductivity sheath (ceramic > stainless steel > copper); insulate the sheath outside.
Stem effect The thermocouple passes through the furnace wall; wall heat loss cools the TC, so the reading is 10–50°C low (depending on wall thickness). Mount through an insulating boss (ceramic tube, refractory); place the tip in the qualified zone centre (the hottest point).
⚠️ shortimmersionerrorA 1000°C furnace with the TC inserted only 20 mm (6 mm sheath, ratio 3:1, should be ≥10:1). Heat conducts along the sheath to the cold environment, so the TC reads 950°C instead of 1000°C — a 50°C error from installation alone, and calibration is irrelevant. To compensate, the operator turns the controller up…

Protection tubes and thermowells

Type Material Highest temperature Application (when to use)
Open circuit / bare junction No protection (bare TC wire) Depends on TC type Fastest response (0.1–0.5 s), transient measurement; cannot be used in corrosive/dirty environments, brittle.
Bare MI cable Stainless steel sheath, bare end 1100°C(SS316) Fast response (0.5–2 s), robust, clean combustion gas/inert atmosphere; the end is vulnerable to impact.
Grounded MI cable SS sheath, end welded to the tip 1100°C Medium response (2–5 s), most common, the end is protected; possible ground loops (use isolation where there is an issue).
Ungrounded MI cable SS sheath, end and tip isolated 1100°C Slow response (5–15 s), eliminates ground loops; for noisy environments or when several TCs share a controller.
CeramicProtection tube Alumina, mullite, silicon carbide 1700°C(alumina) High-temperature applications; essential for noble-metal TCs (contamination protection); poor thermal shock (ramp slowly).
Thermowell (process connection) bar stockmachining(SS, Inconel, Hastelloy) Depends on material For permanent installation in pressure vessels/pipelines, allowing TC replacement without shutdown; designed to ASME PTC 19.3 (pressure class).

Installation method (furnace wall, most common)

  1. Drill a hole in the refractory layer: hole diameter = sheath outside diameter + 2 mm clearance, drilled perpendicular to the wall (never at an angle).
  2. Fit a ceramic guide tube: insert an alumina tube through the wall (protects the TC from refractory abrasion and air ingress); tube inside diameter = sheath outside diameter + 1 mm.
  3. Insert the thermocouple: push the TC through the guide tube to the required immersion depth and mark the sheath at the outer wall (reference point).
  4. Fit the compression fitting: slip the sheath onto the furnace outlet, tighten the nut to seal (prevents hot gas escape); do not over-tighten (crushing the sheath, breaking wires).
  5. Seal the annular gap around the TC: pack ceramic fibre rope at the outer wall to prevent heat loss and air leakage.
  6. Connect the extension wire: use temperature-rated terminal blocks, check polarity, apply labels.
⚠️ Do not over-tighten the compression fitting compression ferrule” hand-tight plus 1/4 turn” seals it. Over-tightening crushes the sheath (especially MI cable with mineral powder); crushing causes the wire to short out and the sensor to fail. If it still leaks after correct tightening, replace the compression ferrule (a consumable); do not apply more force.

Response time vs. protection trade-off

Speed and maximum protection cannot both be had; choose by application priority:

Priority Configuration response time Service life
Speed (transient testing, TUS tuning checks) bare end, 1.5mm TC 0.1–0.5 seconds Short (days to weeks), oxidation/mechanical damage, disposable.
Balanced (process control) Grounded MI cable, 6mm 2–5 seconds Medium (6–12 months), suits most industry.
durability(continuousoperate) TC in a ceramic tube + thermowell 20–60 seconds Long (2–5 years), maximum protection; for processes that must not shut down to change sensors.

Instrument class (temperature accuracy)

Class Sensoraccuracy controller/recorderaccuracy System accuracy (overall) application
Class 0 ±0.25% or ±0.5°C ±0.15% or ±0.3°C ±0.4% or ±0.8°C Most stringent, rare, special processes (single-crystal blades, additive manufacturing).
Class 1 ±0.4% or ±0.75°C ±0.3% or ±0.5°C ±0.7% or ±1.25°C High accuracy (precipitation hardening, nickel-alloy solution treatment).
Class 2 ±0.75% or ±1.5°C ±0.5% or ±1.0°C ±1.25% or ±2.5°C Most common; general aerospace heat treatment (quench & temper, normalising, annealing).
Class 3 ±1.5% or ±3.0°C ±1.0% or ±2.0°C ±2.5% or ±5.0°C Non-critical (stress relief, preheating, general furnaces).
Class 4 ±3.0% or ±6.0°C ±2.0% or ±4.0°C ±5.0% or ±10.0°C Loosest; cleaning furnaces, burn-off furnaces, non-metallurgical processes.
Note: system accuracy = sqrt(Sensor^2 + controller^2) – errors combine by root-sum-square (RSS), not by linear addition.

TUS (temperature uniformity survey)

Purpose:Verify the furnace qualified zone (working area) temperature uniformity is within plus/minus 5 C (Class 2) or tighter. Perform before production and at intervals (quarterly/half-yearly/yearly, depending on furnace type and stability).

Furnace working volume Minimum TUS sensor count Sensor placement
<3 ft³ (85 litres) 3 Centre + 2 corners of the qualified zone
3–27 ft³ (85–765 litres) 6 Centre + 5 surrounding points (walls, top, bottom)
>27 ft³ (>765 litres) 9 Centre + 8 points (front/rear/left/right/top/bottom perimeter + middle)

TUS procedure (Class 2 example)

  1. Calibrate the reference instruments: all TUS sensors calibrated within 6 months (Class 2 system accuracy plus/minus 2.5 C), with certificates on file.
  2. Mount sensors on the load: position per AMS 2750, using a representative heat load (steel blocks, similar to production parts).
  3. Connect a data recorder: multi-channel recorder (at least 9 channels for large furnaces), sampling rate ≥1 per minute.
  4. Heat to the setpoint: typical test temperatures 150°C, 500°C, 850°C (low/medium/high), stable (all sensors within ±5°C of setpoint for 15 minutes).
  5. Record 30 minutes: once stable, compute delta T = (Tmax – Tmin) at each time point; if delta T over the full 30 minutes is less than or equal to plus/minus 5 C (Class 2), uniformity PASSES.
  6. Repeat at all test temperatures: check low/medium/high; some specifications add test points (e.g. 200°C, 650°C, 1050°C).
  7. Record the results: the TUS report includes date, furnace ID, load configuration, sensor calibration date, time-temperature curve, maximum delta T, pass/fail, signed off by a qualified engineer.
  8. Definition of qualified zone: the volume where uniformity is acceptable; marked on the furnace (physical label / painted working grid), the load must sit within this zone.
⚠️ TUS failure If TUS fails (delta T > plus/minus 5 C) possible causes: fan fault / low airflow (poor circulation); heating-element burnout (uneven heating); door-seal leak (cold air into corners); TC position shifted (control sensor no longer representative); load blocking airflow (obstructed flow). Fix the cause and re-test; do not produce until passing, downtime costs thousands of pounds per day.

SAT(systemaccuracytesting)

Purpose:Verify that the control instruments (sensor + controller + recorder) read the correct temperature within system accuracy, using independent verification against a calibrated reference instrument.

frequency method
Quarterly (3 months) or semi-annually (6 months) Depends on instrument class and furnace type (see AMS 2750 Table 6)
After sensor replacement A SAT is required within 7 days of replacing the control sensor
After controller replacement/repair A SAT should be performed before resuming production

SAT steps

  1. Insertion-calibration test TC: place within 75 mm of the control sensor (same temperature zone); test TC calibrated within 12 months, plus/minus 1.0 C accuracy (Class 2).
  2. Connect a calibration indicator: a portable temperature indicator (calibration certificate ≤12 months), independent of the furnace controller.
  3. Heat to the test temperature: typically 150°C, 500°C, 850°C, stable for 15 minutes.
  4. Compare readings: simultaneously record the control sensor (controller display) and the test sensor (portable indicator).
  5. Compute system error: error = (control temperature) – (test temperature); must be within system accuracy (Class 2 plus/minus 2.5 C).
  6. Record: the SAT form includes date, furnace, instruments used, test temperatures, errors and pass/fail, filed in the furnace record.

Dingluxuan Sensor Technology

Design, manufacturing and sales of temperature, pressure and humidity sensors, instruments and ceramic products — delivering integrated temperature-measurement and control solutions.

Products

Products

Applications

Applications

Contact Us

Contact
No. 828 Xueyuan Road, Bacheng Town, Kunshan City, Jiangsu, China
Tel
0512-57892201
Mobile: 13451695112
Email: yanghui@dingluxuan.com
Copyright©2017-2026  Suzhou Dingluxuan Sensor Technology Co., Ltd.   苏ICP备17068273号-1
中文EN
中文EN